  "title": "Physical Env Feeding & Passaging",
  "status": "Draft protocol · physical validation pending",
  "scope": "One maintenance cycle for the assigned iPSC cultures, followed by one approved passage from a source well into a new coated six-well plate, with Squid imaging and recorded reviewer decisions.",
  "boundary": "This page defines the workflow for review; it does not connect to Trillo or Squid, start a run, or certify a physical procedure.",
  "roles": [
    {
      "title": "Trillo",
      "text": "Execute approved liquid-handling methods with the configured P1000, tips, labware, deck positions and calibrated movements."
    },
    {
      "title": "Squid microscope",
      "text": "Acquire real images using the selected plate map, objective, fields, channels and focus method; save image files with plate, well and run identities."
    },
    {
      "title": "Operator and scientist",
      "text": "Stage reagents and plates, perform uncommissioned transfers and tapping/collection, review images, and authorize continuation, extension or stop."
    }
  ],
  "decisions": [
    {
      "title": "Choose one approved biological protocol",
      "bullets": [
        "The scientist-session workflow uses visual cell coverage to choose 2 / 3 / 4 mL feeds and describes a 1–4 minute release incubation.",
        "The pinned repository workflow uses a fixed 2 mL feed and a 5-minute incubation.",
        "The manufacturer describes media- and cell-line-dependent timing, including 6–8 minutes for its mTeSR Plus example.",
        "The scientist must select the applicable SOP, medium, matrix, incubation time and extension rule; do not combine these recipes automatically."
      ],
      "id": "decisions-1"
    },
    {
      "title": "Approve feed cadence and recovery timing",
      "bullets": [
        "The pinned repository baseline schedules maintenance every 48 hours, preferably at 09:00 with a four-hour tolerance, and a next-morning feed after recovery in ROCK inhibitor with at least ten hours since recovery; confirm the appropriate schedule and overrides for these actual cultures."
      ],
      "id": "decisions-2"
    },
    {
      "title": "Approve Trillo settings for that recipe",
      "bullets": [
        "Record source/destination wells, deck and lid-park locations, gripper/tool identity, liquid levels, aspiration position, tilt method, dispense position, flow rates and tip policy.",
        "Confirm well capacity and validate any 4 mL feed against the actual method limits.",
        "Simulator residuals and nominal trajectories are not physical calibration."
      ],
      "id": "decisions-3"
    },
    {
      "title": "Approve the split and dilution plan",
      "bullets": [
        "Record source volume, destination-well count, target volume per well, added medium and expected leftovers.",
        "For the session-notes six-well / 2 mL target, account explicitly for 12 mL of final suspension; adding 1 mL before collection alone does not supply this volume."
      ],
      "id": "decisions-4"
    },
    {
      "title": "Select the physical transfer route",
      "bullets": [
        "Use the commissioned long-gripper Squid handoff only after confirming it for the present robot, plate and setup; otherwise assign an operator carry.",
        "Incubator transfers, lid handling, tapping and collection must each have an assigned actor and a completion record."
      ],
      "id": "decisions-5"
    },
    {
      "title": "Approve the Squid imaging recipe",
      "bullets": [
        "Select the actual objective, well map, repeatable fields, illumination/channel, exposure, focus method, output location and file naming.",
        "The pinned lab workflow proposes 4× imaging.",
        "Confirm target-unit support before use: repository documentation reports a generic autofocus stub and recommends checking the available laser-focus path."
      ],
      "id": "decisions-6"
    }
  ],
  "setup": [
    {
      "actor": "Scientist",
      "title": "Assign cultures and approve the run",
      "evidence": "Run manifest with plate/well identities, due times, approved recipe and reviewer.",
      "id": "setup-1",
      "bullets": [
        "List every plate, cell line and assigned well.",
        "Record passage number, last feed, last passage and recovery treatment.",
        "Choose the approved SOP revision, cultures due for feeding, source well and new destination wells.",
        "Assign the reviewer and approve the run manifest."
      ]
    },
    {
      "actor": "Operator",
      "title": "Prepare the physical workcell",
      "evidence": "Deck/reagent inventory and destination-coating record.",
      "id": "setup-2",
      "bullets": [
        "Match tools, tips, reservoirs, waste and plates to the approved deck map.",
        "Stage the identified medium, PBS and ReLeSR at the SOP-required condition.",
        "Confirm destination coating, collection tube and specified collection pipette.",
        "Check reagent volumes, waste capacity and lid-park locations."
      ]
    },
    {
      "actor": "Operator + instrument owner",
      "title": "Confirm motion and imaging readiness",
      "evidence": "Setup review and instrument-readiness records; approved movement and imaging profiles.",
      "id": "setup-3",
      "bullets": [
        "Confirm the current Trillo calibration and the approved movement profile.",
        "Choose the commissioned robot handoff or assign an operator carry.",
        "Verify Squid loading orientation, plate lock, chamber lid and objective clearance.",
        "Confirm the imaging recipe and writable output location.",
        "Assign a person to watch physical motion."
      ]
    }
  ],
  "feeding": [
    {
      "actor": "Operator or commissioned handoff",
      "title": "Move a due plate to Squid",
      "evidence": "Plate identity, source/destination, transfer completion and timestamps.",
      "id": "feeding-1",
      "bullets": [
        "Identify the due plate before removing it from its incubator.",
        "Record the departure time.",
        "Carry or hand off the plate using the approved route.",
        "Load Squid in the approved orientation."
      ]
    },
    {
      "actor": "Squid + scientist",
      "title": "Image and review the wells",
      "evidence": "Real image files, acquisition metadata, well-level assessment and decision.",
      "id": "feeding-2",
      "bullets": [
        "Acquire the approved fields for each assigned well.",
        "Confirm that the real image files are readable and in focus.",
        "Have the scientist assess cell coverage, morphology and debris.",
        "Record proceed or hold and the feed amount for each well."
      ]
    },
    {
      "actor": "Trillo + operator where required",
      "title": "Replace medium one well at a time",
      "evidence": "Executed method and parameter revision, actual command outcomes, selected volume, tips and any exceptions.",
      "id": "feeding-3",
      "bullets": [
        "Identify the plate and well on the Trillo deck; manage the lid using its assigned park.",
        "Remove spent medium using the validated aspiration and tilt method without touching colonies.",
        "Add the approved fresh-medium amount down the well wall.",
        "Finish removal and replacement for one well before moving to the next."
      ]
    },
    {
      "actor": "Operator or commissioned handoff",
      "title": "Close and return the plate",
      "evidence": "Incubator-return receipt, times and exception review.",
      "id": "feeding-4",
      "bullets": [
        "Replace the lid and verify the plate identity.",
        "Return the plate to its assigned incubator within the approved timing limits.",
        "Record the return time.",
        "Review unexpected liquid loss, contact or handling errors before accepting completion."
      ]
    },
    {
      "actor": "Scientist",
      "title": "Record maintenance and the next feed",
      "evidence": "Culture maintenance record and next-feed obligation.",
      "id": "feeding-5",
      "bullets": [
        "Record each fed well and the fresh-medium amount added.",
        "Record any held wells and the scientist’s disposition.",
        "Set the next feed time using the selected SOP.",
        "Apply the approved recovery-treatment override where relevant."
      ]
    }
  ],
  "passaging": [
    {
      "actor": "Scientist + Squid",
      "title": "Confirm the source and capture before-images",
      "evidence": "Source-well before-images, eligibility assessment and signed reviewer decision.",
      "id": "passaging-1",
      "bullets": [
        "Identify the source well and its cell line.",
        "Acquire the approved before-treatment fields on Squid.",
        "Have the scientist assess cell coverage and purity.",
        "Record passage approval or hold, with links to the images."
      ]
    },
    {
      "actor": "Operator + Trillo",
      "title": "Prepare the new plate",
      "evidence": "Destination identity, coating record, completed preparation and per-well starting volumes.",
      "id": "passaging-2",
      "bullets": [
        "Verify the coated destination plate and all assigned wells.",
        "Remove coating liquid using the approved method.",
        "Add destination medium only if specified by the selected recipe.",
        "Keep preparation contiguous so the coating does not dry."
      ]
    },
    {
      "actor": "Trillo",
      "title": "Wash and apply ReLeSR",
      "evidence": "Timed method trace, actual liquid-handling outcomes and recorded deviations; the session notes specify one wash and a 1 mL ReLeSR addition.",
      "id": "passaging-3",
      "bullets": [
        "Remove spent medium from the assigned source well.",
        "Perform the selected SOP’s PBS wash and remove it without an unrelated pause.",
        "Add and remove the approved ReLeSR amount, leaving the intended thin film.",
        "Record the command trace and actual deviations; the session-notes route specifies one wash and 1 mL ReLeSR."
      ]
    },
    {
      "actor": "Operator + incubator",
      "title": "Incubate using the approved time",
      "evidence": "Selected SOP timing, incubation record and transfer times.",
      "id": "passaging-4",
      "bullets": [
        "Replace the lid and transfer the source using the approved route.",
        "Start the incubation timer at the point defined by the selected SOP.",
        "Use the approved condition and treatment time.",
        "Record actual incubation start/end times separately from transport."
      ]
    },
    {
      "actor": "Squid + scientist",
      "title": "Compare before/after images and decide",
      "evidence": "Matched image pairs, actual accumulated treatment time and reviewer decision with rationale.",
      "id": "passaging-5",
      "bullets": [
        "Acquire the same source fields after treatment.",
        "Have the scientist compare the before/after images.",
        "Record proceed, extend or stop, with a rationale.",
        "For extend: follow the approved extension limit, then reimage and rereview.",
        "For stop: do not collect or split; record the scientist’s disposition."
      ]
    },
    {
      "actor": "Trillo + operator",
      "title": "Add medium, loosen and collect the cells",
      "evidence": "Added-medium record, tapping/collection completion, collection tool, tube identity and any observed splash or loss.",
      "id": "passaging-6",
      "bullets": [
        "Proceed only after the reviewer authorizes collection.",
        "Add the medium amount specified by the selected SOP.",
        "For the session-notes route, gently tap the plate side and check for splash.",
        "Collect into the identified tube with the specified 5 mL serological pipette.",
        "Keep culture lines separate; record any splash, loss or interrupted collection."
      ]
    },
    {
      "actor": "Operator",
      "title": "Mix and prepare the reviewed dilution",
      "evidence": "Mixing and dilution record, collected/added/final volumes, and operator confirmation.",
      "id": "passaging-7",
      "bullets": [
        "For the session-notes route, pipette the collected cells up and down once.",
        "Add medium according to the approved dilution plan.",
        "Record collected volume, added volume, final volume and any leftovers.",
        "Maintain the mixture during distribution as the selected SOP requires."
      ]
    },
    {
      "actor": "Assigned operator or validated Trillo method",
      "title": "Seed the new wells and return plates",
      "evidence": "Per-well seeding volumes, leftover/waste account, settle/return records and any interrupted work.",
      "id": "passaging-8",
      "bullets": [
        "Use an operator unless the robot’s tube-to-plate method is commissioned.",
        "Deliver the approved amount to every assigned destination well.",
        "Perform the selected SOP’s settling/distribution handling.",
        "Relid and return source and destination plates to their assigned locations.",
        "Record seeding volumes, leftovers, waste and transfer completion."
      ]
    },
    {
      "actor": "Scientist",
      "title": "Register the new culture and follow-up",
      "evidence": "Parent/new-plate relationship, passage record and future feed/imaging obligations.",
      "id": "passaging-9",
      "bullets": [
        "Link the new plate to its source culture and well.",
        "Increment and record the passage number.",
        "Record the split and any exceptions.",
        "Schedule the next feed and required follow-up imaging under the cell-line SOP."
      ]
    }
  ],
  "imaging": [
    {
      "title": "Match identity and fields",
      "bullets": [
        "Each acquisition must identify the run, plate, well, time point, objective, channel, exposure, focus method and field positions; compare the same source fields before and after release."
      ],
      "id": "imaging-1"
    },
    {
      "title": "Verify image delivery",
      "bullets": [
        "Confirm acquisition completion and that the actual files can be opened before accepting an imaging step; a command acknowledgement, filename alone or simulated receipt is insufficient."
      ],
      "id": "imaging-2"
    },
    {
      "title": "Separate acquisition from interpretation",
      "bullets": [
        "An image does not itself approve passage: preserve the scientist’s assessment and proceed/extend/stop decision linked to those files."
      ],
      "id": "imaging-3"
    }
  ],
  "completion": [
    "Every assigned feed is completed or explicitly held with a recorded disposition; plates are returned and their next due times are recorded.",
    "Passage follows an approved image-based decision, with either documented transfer into the new plate or a documented scientist-directed stop.",
    "Real Squid images, reviewer decisions, executed Trillo method/parameter identities and operator transfer records are linked to the same run and wells.",
    "Tools, tips, lids, plate locations, reagent use, waste, destination volumes and leftovers are accounted for, with exceptions reviewed rather than silently retried.",
    "This draft becomes a runnable protocol only after the unresolved recipe, calibration, imaging and handoff settings are approved and the physical methods are validated."
  ],
  "sources": [
    {
      "title": "Scientist session: feeding and selective passaging notes",
      "url": "ipsc.html#evidence",
      "note": "Session-derived handling choices; not a signed physical SOP."
    },
    {
      "title": "Pinned automation maintenance protocol",
      "url": "https://github.com/engrambio/automation/blob/5cdc95435a23e3acc0d2ffc3e9ce922d58cb974b/tools/workcell/protocols/library/ipsc-maintenance.json",
      "note": "Repository baseline: fixed-volume maintenance and instrument/handoff steps."
    },
    {
      "title": "Pinned automation passage protocol",
      "url": "https://github.com/engrambio/automation/blob/5cdc95435a23e3acc0d2ffc3e9ce922d58cb974b/tools/workcell/protocols/library/passage.json",
      "note": "Repository baseline: ReLeSR workflow, review and parent/new-plate registration."
    },
    {
      "title": "Squid instrument and handoff documentation",
      "url": "https://github.com/engrambio/automation/blob/5cdc95435a23e3acc0d2ffc3e9ce922d58cb974b/docs/instruments/cephla-squid.md",
      "note": "Verify against the target microscope and commissioned robot setup."
    },
    {
      "title": "Squid client limitations",
      "url": "https://github.com/engrambio/automation/blob/5cdc95435a23e3acc0d2ffc3e9ce922d58cb974b/tools/squid/README.md",
      "note": "Command support and reported focus limitations."
    },
    {
      "title": "STEMCELL ReLeSR protocol",
      "url": "https://www.stemcell.com/enzyme-free-passaging-of-hpscs-using-relesr.html",
      "note": "Manufacturer cross-check: media/cell-line-dependent incubation; does not approve the local adaptations."
    }
  ],
  "gates_file": "physical-env-gates.json"
}